Jual: Autoclave ES-315
Reliable & Speedy Sterilization

A top-open lid, which can be operated easily with a single hand or foot, is provided with the SX autoclaves. To open the lid, the lid lock is released by stepping on the foot pedal locates at the bottom while pressing the lid down. The lid can be opened with minimal effort owing to mechanical assistance. In addition, the lid can be closed easily just by pressing it down.
User-friendly Memory Function

Once turned on the system automatically sets and displays a pre-set value. Operation is done simply by pressing the start switch. ES-315/ ES-215 enables the set temperature and time for each mode program to be stored in its memory.

Model ES-315
Operating temperature range
( during sterilizing) 105 - 132° C
Operating pressure range 0 - 186kPa
Maximum operating pressure 216kPa
Temperature control Digital, microprocessor controlled
Temperature display / display range Digital / -15 - 180° C
Pressure display / display range Analog / 0 - 400kPa
Heat source 2.0kW electric heater
Timer control
Digital, microprocessor controlled
Timer display Digital
Timer
display range Sterilizing mode: 1 - 240min. or continuous
Sterilizing-warming mode: 1 - 240min. ( sterilizing) / 1 - 8hours ( warming)
Heating mode: 1 - 8hours
Start-timer setting range 1 - 99hours in each operation mode
Memory function A set of temperature and time setting in each operation mode
Safety devices • Inside the chamber overheat prevention, • Outer wall of the chamber overheat prevention, • Overpressure prevention, • Temperature sensor disconnection prevention, • Empty heating prevention, • Leakage breaker, • Safety valve
Malfunction prevention devices • Low water level detection, • Exhaust valve knob open/ close detection,
• Insufficient sterilization detection, • Chamber lid open/ close detection
Leakage
breaker Rated breaking current: 30A ( 120V) , 15A ( 220 / 230 / 240V)
Rated sensed current: 30mA ( 120V) , 10mA ( 220 / 230 / 240V)
Protection type Class 1 equipment
Operating
environment Ambient temperature: 5 - 35° C
Relative humidity: 30 - 85%
Atmospheric pressure: 700 - 1060hPa
Gradient: Within 2°
Dimensions Æ 325 x 740mm
Usable dimension Æ 312 x 635mm
Capacity of chamber 53
Chamber material SUS304
Capacity / material of exhasut bottle 3 / polyethylene
Type / material of lid gasket Accommodating internal pressure / silicone rubber
Dimensions of main unit ( mm) 490W x 560D x 1090H ( Height from floor to control panel: 875)
Net weight 80kg
Rated Voltage AC 120 / 220 / 230 / 240V
Power input 20A ( 120V) , 10A ( 220V) , 9A ( 230 / 240V)
Power supply requirements Single-phase 220 / 230 / 240V AC ( 50 / 60Hz) 15A or above
Power Comsumption ( calorific value) 2kW ( 1720kcal/ h)
Accessories • Stainless steel basket with slatted bottom plate 1, • Water level sensor 1, • Inspection sheet 1, • Warranty card 1, • Customer card 1, • Operator' s manual 1, • Clear plastic accessory case 1, • Attachment screw for clear plastic accessory case 1
Stainless basket 2 ( Æ 300 x 182mm)
Negara Asal:Jepang